| Product Detail: | 
Vibration Sample Magnetometer
Main Function of Vibration Sample Magnetometer
The VSM (also named M-H CURVE Hysteresis Graph TEST SYSTEM ) can be used for the measurement of the basic magnetic properties of the magnetic materials. Such as the Hysteresis loops of the materials, Magnetization curve, Warming Curve , Warming /Cooling curve, Cooling curve and the changes of the temperature as time goes away , then we will get some Magnetic parameters such as the Saturated strength of the magnetization, the left strength of the magnetization , Coercive force, the Max of Magnetic energy product,the Curie temperature and the Magnetic conductivity(the Initial Magnetic conductivity is included ). It can also measure the magnetic materials such as the powder, the particles, the films, the liquid and the massive.
The VSM can measure the changes of magnetic with the changes of temperature from -196℃ to 900℃.
Main Specifications of Vibration Sample Magnetometer
- Measure the scope of the magnetic moment: 10-2emu—300emu(Sensitivity:5×10-5emu).
 - Relative accuracy (30emu ): better than ±1%
 - Repeatability (30emu): better than ±1%.
 - Stability (30emu) : Preheating 24 hours , 24 hours of continuous work superior ±1%.
 - Temperature range : from -196℃ to 900℃.
 - The magnetic Pole pitch for a fixed spacing is 40 mm , the pole face diameter is 60 mm.
 - Magnetic field: supplied by the electromagnet, from 0 to 3.5T .
 -  Material to be measured:
• Diamagnetic, Paramagnetic, Ferromagnetic, Ferrimagnetic, Antiferromagnetic materials and
Anisotropic materials
• Particulate and continuous magnetic recording materials and GMR, CMR, exchange biased and spinvalve materials
• Magnetic-optical materials
• Bulk materials, powders, thin films, single crystals, and liquids are readily accommodated 
Constitution of the VSM.
No.   | 
    VSM
      550~100   | 
  
Electromagnet  | 
    √  | 
  
Steady
      Flow
      Power  | 
    √  | 
  
Vibrating
      head、 
      Vibrating
      frame   | 
    √  | 
  
Vibrating
      staff 、 
      Sample
      room   | 
    √  | 
  
Vibrating Source   | 
    √  | 
  
Lock
      in Amplifier   | 
    √  | 
  
Gauss meter  | 
    √  | 
  
Detection coil  | 
    √  | 
  
Computer  | 
    √  | 
  
Printer   | 
    √  | 
  
VSM can prepare High and Low Temperature Device separately.
Note of the Main Equipment
1. Electromagnet
The electromagnet is of the type adjustable Double-conjugated or the fixed Gap. It will be placed with a 45-degree angle ground .There are several configuration as the following :
No.  | 
      At room    temperature, the strength in the center of the magnetic    field  | 
      At high and low    temperature the strength in the center of the magnetic    field  | 
      The manner of cooling  | 
    
VSM-550  | 
      3.8  | 
      3.4  | 
      Cool with water  | 
    
VSM-400  | 
      3.3  | 
      3.0  | 
      Cool with water  | 
    
VSM-380  | 
      3.0  | 
      2.7  | 
      Cool with water  | 
    
VSM-300  | 
      2.6  | 
      2.3  | 
      Cool with water  | 
    
VSM-250  | 
      2.4  | 
      2.1  | 
      Cool with water  | 
    
VSM-220  | 
      2.0  | 
      1.7  | 
      Cool with water  | 
    
VSM-175  | 
      1.8  | 
      1.5  | 
      Cool with water  | 
    
VSM-130  | 
      1.5  | 
      1.2  | 
      Cool by itself  | 
    
VSM-100  | 
      1.0  | 
      0.8  | 
      Cool by itself  | 
    
2. Steady Flow Power
The  power is the automatic conversion DC Power which is of high stability,  steady voltage and steady flow. The scope of the power is from 2 to 30KW. In  the state of the steady flow, the current that Steady Flow outputs is continuous  adjustable in the rated scope. 
  Main  features 
(1) The rated output voltage: the rated output power from  0 to 12.0KW
(2) Protection: Lack item protection, Overflow protection and  Automatic short-circuit protection
Specifications 
(1) Power output is for the steady flow: the scope of the  current value is from 0 - rated continuous-tuning.
(2) Display mode: The ampere meter is Four-semi-LED  Digital.
(3)  The accuracy  of  Current instructions : ±(1%+2 words) 
(4) When the load is the electromagnet, and output  current is more than half the maximum current, the current stability the power  outputs is better than 5 × 10-4
(5) Time for work:   work for uninterrupted 8 hours. (Environmental temperature 20±5℃)
(6) Input voltage:         AC380V±10%
(7) Input frequency:       50Hz
3. Vibration System
It includes vibration head, Vibrating frame, sample room and the detection coils.
4. Magnetic measurement unit
- The scope  is among “300emu”、 “150emu” 、“80emu” 、“40emu”、“30emu” 、“15emu” 、“8emu”、“4emu”、“3emu”、“1.5emu”、
“800memu”、“400memu”、“300memu”、
“150memu”、“80memu”、“40memu”、“30memu” and “15memu”.We can change it among the figures. - There are several scopes of the strength of the magnetic field. 0.5kOe”、“1kOe”、“2kOe”、“4kOe”、“8kOe”、“16kOe” and “32 kOe” . It can be shown throngh the Four semi-LED digital control. The resolution is 0.1mT. The relative accuracy is superior ±1%.
 - The output power of the Vibration source is 185Hz. Frequency stability  is superior 10-5.   The output power is more than 50W.
 
5. The computers of the brand of Lenovo
PDC E2140, Memory :1G. Hard disk : 160G. Monitor : 17-inch LCD
6. HP Printer : HP –1018
7. High temperature furnace and Temperature Control Device
The heating power is 100W, The scope of the furnace temperature is from the room temperature to 900 ℃.
It is shown through Four semi-LED digital control .The resolution is 0.1 ℃.
8. Hypothermia Dewar and Temperature Control Device
- The control scope of the temperature of the Samples room is between 77K to 373K.
 - It is shown through Four semi-LED digital control .The resolution is 0.1K.
 
NdFeB production line